Maintenance
Source: Intuition-6 Instruction Manual Rev. H, Page: 102 Regular maintenance ensures accurate measurements and reliable control. The primary maintenance tasks involve sensor cleaning and calibration.Maintenance Schedule
Electrode Cleaning
Use these procedures to restore sensor accuracy and prevent fouling-related drift.Contacting Conductivity Electrodes
Graphite Electrodes (Cooling Tower):- Remove from process
- Rinse with clean water
- Brush gently with soft brush to remove deposits
- For heavy scaling, soak in dilute acid (10% HCl) for 15 minutes
- Rinse thoroughly before reinstalling
- Remove from process
- Rinse with clean water
- For scale, soak in dilute phosphoric acid
- For oil/grease, clean with mild detergent
- Rinse and reinstall
Electrodeless Conductivity Sensors
- Remove from process
- Rinse with clean water
- Use soft cloth to wipe sensing surfaces
- For stubborn deposits, use appropriate chemical cleaner based on deposit type
- Rinse thoroughly and reinstall
pH Electrodes
Routine Cleaning:- Remove from process
- Rinse with clean water
- Soak in pH electrode cleaning solution for 15 minutes
- Rinse with clean water
- Store in pH 4 buffer or KCl storage solution if not reinstalling immediately
- Soak in 0.1 M HCl or pepsin solution
- Follow with standard cleaning procedure
- Soak in dilute EDTA solution
- Rinse thoroughly
Never wipe pH glass bulb with dry cloth—this generates static charge that affects readings. Always rinse with water.
ORP Electrodes
- Remove from process
- Rinse with clean water
- Clean platinum surface with soft cloth
- For stubborn deposits, polish with aluminum oxide slurry
- Rinse and reinstall
Disinfection Sensors
Membrane-Style Sensors:- Remove flow cell from process (close isolation valves)
- Remove sensor from flow cell
- Rinse sensing surface with clean water
- Check membrane for tears or deposits
- Replace membrane cap if damaged or per schedule
- Follow manufacturer-specific cleaning procedures
- May require electrolyte replacement
Sensor Evaluation
Use these checks to determine whether sensors need cleaning or replacement.Evaluating Conductivity Electrodes
Contacting Conductivity:- Measure resistance between electrode terminals with multimeter
- Compare to expected values:
- 0.01 cell: Higher resistance
- 10.0 cell: Lower resistance
- Clean if readings are inconsistent
- Check Installation Factor after calibration
- Should be between 0.5 and 1.5
- Values outside range indicate coating or positioning issues
Evaluating pH/ORP Electrodes
pH Electrode Tests:- Calibrate with two buffers
- Check slope value:
- Good: 54-62 mV/pH
- Marginal: 48-54 mV/pH (consider replacement)
- Failed: Below 48 mV/pH (replace)
- Check response time (should stabilize within 60 seconds)
- Place in ORP standard solution
- Compare reading to expected value
- Check response time
- Replace if response is slow or reading is significantly off
Replacing the Powered Relay Fuse
The six powered relays share a common 6.3 A fuse.1
Disconnect power
Turn off the power switch AND disconnect power at the source.
2
Remove front panel
Remove the four screws and carefully open the front panel.
3
Locate fuse
The fuse is located on the power/relay board.
4
Remove old fuse
Gently pull the fuse from its holder.
5
Inspect
A blown fuse will have a broken wire or discolored glass.
6
Install new fuse
Insert replacement fuse: 6.3 A, 5 x 20 mm, 250V fast-acting.
7
Reassemble
Close front panel, reinstall screws, reconnect power.
If the fuse blows repeatedly, check for short circuits in connected equipment. Total relay load must not exceed 6 A.
Battery Replacement
The BR2032 lithium coin cell maintains the real-time clock during power outages. Symptoms of Low Battery:- Date/time resets after power cycle
- Time drifts significantly
- Disconnect power
- Open front panel
- Locate battery on main controller board
- Remove old battery, note polarity
- Insert new BR2032 battery
- Reassemble and power on
- Set correct date and time
Calibration Frequency Guidelines
Preventive Maintenance Checklist
Use these checklists to keep the controller and sensors in good working order.Monthly
- Check all sensor readings for reasonableness
- Verify flow through sample lines
- Clean sensors if needed
- Review alarm history
- Check chemical tank levels
- Verify relay operation (put in Hand mode briefly)
Quarterly
- Calibrate conductivity sensors
- Clean and calibrate pH sensors
- Calibrate ORP sensors
- Verify flow meter calibration
- Check all wiring connections for tightness
- Inspect enclosure seals
Annually
- Calibrate temperature sensors
- Calibrate 4-20 mA transmitters
- Inspect and test fuse
- Check battery voltage
- Verify all relay operations under load
- Review and update control settings
- Export configuration to USB for backup
Spare Parts
Keep these items on hand for quick repairs:Sources
Intuition-6 Instruction Manual Rev. H
Page 102 • February 2024