Startup
Source: DICE IOM Manual V12E , Pages: 18–23
The module is internally pressurized and handles hazardous chemicals. Use appropriate PPE: protective gloves, glasses, and clothing. Refer to the MSDS for all fluids used.
If valves were factory preset (labeled with target pressure), skip the adjustment procedures and proceed to calibration.
Pump Priming
The bleed valve (32) on the left side of the DICE module helps evacuate trapped gas between the pump outlet and back pressure valve.
Bleed valve location on DICE module
Start the pump
Begin pump cycling while observing for flow.
Open the bleed valve slightly
Partially unscrew the bleed valve (32) until fluid begins passing through. Only slight opening is needed—do not fully unscrew.
Close the bleed valve
Once fluid flows, close the bleed valve and proceed with setup.
The bleed valve can also assist if the pump loses prime during normal operation.
Pressure Relief Valve Adjustment
Pressure relief valve adjustment components
Close all ball valves
Set all three (DS) or four (DM) ball valves to the closed position.
Pre-load the back pressure adjuster
Screw the back pressure adjustment screw (19’) completely in (clockwise).
Start the pump
Run the pump until fluid passes through the pressure relief valve into the evacuation port.
Set the pressure relief set point
Use adjustment screw (19) to obtain the desired relief pressure. Clockwise increases the set pressure.
Lock the setting
Tighten counter nut (20). Max torque: 35 in-lb .
Depressurize
Stop the pump and slowly open the evacuation ball valve to release system pressure.
Back Pressure Valve Adjustment
Back pressure valve adjustment components
Set valve positions
Close the evacuation and calibration column valves. Open the injection valve.
Start the pump
Run the pump until fluid passes through the back pressure valve.
Set the back pressure set point
Initially unscrew adjustment screw (19’), then adjust to desired pressure. Clockwise increases the set pressure.
Lock the setting
Tighten counter nut (20’). Max torque: 35 in-lb .
Set back pressure to 20-30 psi above process pressure for optimal check valve operation and siphon prevention.
Adjustment components reference
Dosing Pump Calibration
Complete back pressure and pressure relief valve adjustment before calibrating the pump.
DICE DS Calibration
DICE DM Calibration
Set valve positions
Close the evacuation valve (3) and injection valve (1). Open the calibration column valve (2).
Perform calibration
Follow your dosing pump manufacturer’s calibration procedure using the calibration column.
Drain the calibration column
Open the evacuation valve (3) to empty the calibration column. Vent the top of the calibration column to enable complete drainage.
Return to injection mode
Close the evacuation valve (3) and calibration column valve (2). Open the injection valve (1).
DS Valve Positions
DICE DM valve positions for calibration
Set initial valve positions
Close the evacuation valve (3) and injection valve (1). Open the calibration column valve (2) and separation valve (4).
Perform calibration
Follow your dosing pump manufacturer’s calibration procedure.
Isolate the chemical tank
Close the chemical tank isolation valve (at tank or on pump suction line).
Switch to drain mode
Close the separation valve (4). Open the injection valve (1).
Drain the calibration column
Open the evacuation valve (3) to drain the calibration cylinder.
Suction lift : Contents drain by gravity to tank below
Flooded suction : Start pump to empty contents into process
Return to injection mode
Close the evacuation valve (3) and calibration column valve (2).
Restore tank isolation
Open the chemical tank isolation valve.
DM Valve Positions
The top of the calibration column must be vented to atmosphere to enable gravity drainage through the evacuation valve.
Post-Startup Verification
Verify pressure
Confirm the pressure indicator reads the back pressure set point.
Check for leaks
Inspect all connections for leaks.
Confirm valve positions
Verify all ball valves are in the correct position for normal operation.
Document set points
Record the set points for future reference.
If the pressure indicator does not show the expected back pressure reading, refer to Troubleshooting .
DICE DS — Suction Lift 1/2” module for suction lift applications
DICE DM — Flooded Suction 1/2” and 1” modules for flooded suction
Sources
DICE IOM Manual V12E Pages 18–23 • July 2022