Startup
Source: DICE IOM Manual V12E, Pages: 18–23If valves were factory preset (labeled with target pressure), skip the adjustment procedures and proceed to calibration.
Pump Priming
The bleed valve (32) on the left side of the DICE module helps evacuate trapped gas between the pump outlet and back pressure valve.
Open the bleed valve slightly
Partially unscrew the bleed valve (32) until fluid begins passing through. Only slight opening is needed—do not fully unscrew.
Pressure Relief Valve Adjustment

Pre-load the back pressure adjuster
Screw the back pressure adjustment screw (19’) completely in (clockwise).
Start the pump
Run the pump until fluid passes through the pressure relief valve into the evacuation port.
Set the pressure relief set point
Use adjustment screw (19) to obtain the desired relief pressure. Clockwise increases the set pressure.
Back Pressure Valve Adjustment

Set the back pressure set point
Initially unscrew adjustment screw (19’), then adjust to desired pressure. Clockwise increases the set pressure.
Adjustment components reference
Adjustment components reference
| Part | Description | Location |
|---|---|---|
| 19 | Pressure relief adjustment screw | Top valve |
| 19’ | Back pressure adjustment screw | Bottom valve |
| 20 | Pressure relief counter nut | Top valve |
| 20’ | Back pressure counter nut | Bottom valve |
Dosing Pump Calibration
Complete back pressure and pressure relief valve adjustment before calibrating the pump.- DICE DS Calibration
- DICE DM Calibration
Set valve positions
Close the evacuation valve (3) and injection valve (1). Open the calibration column valve (2).
Perform calibration
Follow your dosing pump manufacturer’s calibration procedure using the calibration column.
Drain the calibration column
Open the evacuation valve (3) to empty the calibration column. Vent the top of the calibration column to enable complete drainage.
DS Valve Positions
| Mode | Valve 1 (Injection) | Valve 2 (Calibration) | Valve 3 (Evacuation) |
|---|---|---|---|
| Calibration | Closed | Open | Closed |
| Draining | Closed | Closed | Open |
| Normal operation | Open | Closed | Closed |
The top of the calibration column must be vented to atmosphere to enable gravity drainage through the evacuation valve.
Post-Startup Verification
If the pressure indicator does not show the expected back pressure reading, refer to Troubleshooting.
Related Products
DICE DS — Suction Lift
1/2” module for suction lift applications
DICE DM — Flooded Suction
1/2” and 1” modules for flooded suction
Sources
DICE IOM Manual V12E
Pages 18–23 • July 2022
